All Categories
EN

Home > Products > By Application > Hospital lsolated Power System

AIM-M300 series medical intelligent insulation monitor


high integration, 

compact size, 

convenient installation and integrates intelligence,

digitalization and networking in one. 

insulation monitoring of isolation power system in Class 2 medical locations


Email: [email protected]

Mobile: +86 134 8217 7205

Whatsapp/Wechat: +86 134 8217 7205

  • Overview
  • Customer Questions & Answers
  • Contact us

Overview

AIM-M300 series medical intelligent monitoring device adopts advanced microcontroller technology, which has high integration, compact size, convenient installation and integrates intelligence, digitalization and networking in one. 

It is suitable for insulation monitoring of isolation power system in Class 2 medical locations such as operating room and intensive care unit.

Function

 Function features of AIM-M300 series medical intelligent monitoring device 

Functions of real-time monitoring and fault alarming of the grounding insulation resistance, transformer load current and transformer winding temperature of the monitored IT system; 

  Locating current injector for insulation fault location systems; 

Real-time monitor the line disconnection fault, current sensor disconnectiobn,temperature sensor disconnection fault and the functional grounding line disconnection fault of the monitored system, and give the alarm indication within 2s after the fault occurs. when the fault occurs; 

  Relay alarm output, LED alarm indication and other faults indication functions; 

Two kinds of fieldbus communication technology, which are used for centralized alarm and display device, insulation fault locator and upper computer management software communications, and monitoring the operation status of IT system in real time. 

  Events logging function, which can record alarm occurrence time and fault type and is convenient for operation personnel to analyze the operation conditions of system and promptly eliminate the faults.

Technical Parameters

AUX PowerVoltageDC 18…36VTemperature 

monitoring

Thermal resistor 2 Pt100
Power consumption≤6WMeasuring range-50…+200℃
Insulation monitoringResistance measuring range15…999kΩAlarm value range0…+200℃
Response value50…999kΩAlarm outputOutput mode2 Relays
Relative uncertainty±10%,±10KContact ratingAC 250V/3A DC 30V/3A
Response time≤3sEnvironmentOperating temperature-10…+55℃

Dimensions

image

Installation

AIM-M300 series insulation monitor adopts the installation method of the DIN-rail, and the fixation mode is the clip buckle type, as shown in the following figure: